DATA ERRORS
Incorrect display or printout results when the measurement optics are displaced a known distance:
SYMPTOM
CAUSE
ACTION
No movement indicated
Coupler interface
1. 10745: check
on all measurement
card not placing
HP-IB address
axes.
instructions on
coupler backplane.
2. Replace or
troubleshoot
10745, or 1 / 0
from coupler
to controller.
System malfunctions
1. 10760A counter
1. Using logic
in extended reso-
fails to re-
probe, verify
lution
ceive extended
10760A receives
resolution corn-
extended reso-
(e.g., laser
mand or reso-
lution command.
readings are 6 or
lution circuitry
(U47 (9) on 10760
10 times greater
is malfunctioning.
should be a
T T L
or less than known
high if in ex-
measured distance)
2. 10760A resolu-
tended resolution)
tion jumpers are
incorrectly set
2. If command not
received, check
proper command
decoder output
on 10760 with
logic probe.
3. Replace inter-
face card.
4. if command re-
ceived, verify
10760A is
correctly jum-
pered for desired
resolution.
5.
Replace 10760A
card.
No movement
10760A Mal-
If possible,
indicated on
function
substitute
one but not all
10760A from
axes.
other axis.