Trouble Shooting
1. The meter does not enter test mode after
inserting a
test strip.
2. The test does not start after applying the
blood drop.
What It Means
• Test strip is inserted upside down, or
bottom end in, or not fully inserted into the
meter.
• No battery is installed
• Battery is installed incorrectly
Dead battery
• Defective test strip or meter
• Blood or foreign objects put in the test strip
port
• Blood drop is too small
• Defective test strip
• Sample applied after meter turns off
• Defective meter
What To Do
Insert test strip print side up, top end in.
Install battery with (+) facing up.
Replace battery. Reset date and time, if
necessary.
If meter still does not enter test mode,
contact Zoetis Customer Care.
Contact Zoetis Customer Care.
Add blood to the same sample area within
60 seconds, or repeat the test using a new
test strip and a larger blood drop.
Repeat the test using a new test strip.
Wait for drop and strip to appear before
applying blood or control solution.
If problem persists, contact Zoetis
Customer Care.
Contact Zoetis Customer Care.
32