Cleaning the AlphaTrak 3 Meter

Avoid getting dirt, dust, blood, control solution or liquid in the meter test strip and data ports. Clean the outside of the meter using a damp
cloth and:
• Mild detergent/soap and water, or
• 70% isopropyl alcohol (available in the USA)
• A mixture of 1 part household bleach, 9 parts water
IMPORTANT:
Do not immerse the meter in water or other liquid.

Error Codes

Message
What It Means
• The sample is too small
• There may be a problem with the test strip
• Meter error
What To Do
 Use a new test strip and try again with
proper amount of sample
 Perform a control solution test using a
new test strip. If the test results are within
the range printed on the test strip vial,
retest using blood and a new test strip.
 If the control solution result is out of range
or the error reappears, contact Zoetis
Customer Care.