1.800.544.2843
5. With the No. 2 pencil, on the frosted portion make an "X" from corner
to corner.
6. Immerse the frosted end of the slide into the solution. Hold the slide
vertically and center it in the solution.
7. Make sure that model CPX/CPHX models are in Timed or Constant
Sonics Mode, not Degas Mode, then turn ultrasonics On.
The ultrasonics will begin immediately to remove the lead from the slide.
All lead should be removed within 10 seconds. If your unit passes this
test, its ultrasonic cavitation is acceptable.
To ensure consistency from test to test, be sure to repeat
test conditions—use the same solution concentration, liquid level,
temperature, type of pencil, length of degassing, etc.
54 Troubleshooting
www.calcert.com
15
10
20
5
25
0
30
NOTE:
sales@calcert.com