Branson 1800 Operator's Manual page 42

Ultrasonic bath
Hide thumbs Also See for 1800:
Table of Contents

Advertisement

1.800.544.2843
CPXH Temperature Calibration
The CPXH unit temperature measurement is factory calibrated. Use the
following instructions to perform periodic calibrations:
Step
Action
1
The ultrasonic bath liquid may be at room temperature or may
be heated-up to a desired operating temperature (e.g. 40 °C).
2
Press the Fn key 4 times until the Current Temperature icon
(
) starts blinking. Press the
Fn key 2 more times to return to the Ready State.
3
Press the On/Standby key (
4
Simultaneously, press both the On/Standby (
key. Only the bottom left digits and the Current Temperature
icon (
) should turn on.
5
Stir the solution for 15 seconds to ensure thermal uniformity.
6
Wait 2 minutes after turning the unit on before taking
measurements. This allows for the display to be properly
updated.
7
Use the
/
match the actual tank temperature.
8
Press the On/Standby key (
37 Operating Your Ultrasonic Bath
www.calcert.com
15
10
20
5
25
0
30
key to select °F. Press the
) to turn off the unit.
keys to change the display temperature to
) to end calibration.
) and the Fn