Page 2
Warranty Ultrasonic Cleaners, when used in accordance with manufacturer’s instructions and under normal use, are guaranteed for two years after date of shipment. Within the period guaranteed, we will repair or replace free of charge, at our sole discretion, all parts that are defective because of material or workmanship, not including costs for removing or installing parts.
Contents Front Warranty Cover Safety Precautions Introduction PC620 General Information Accessories Unpacking Your Cleaner Installing Your Cleaner Equipment Specifications How Ultrasonic Cleaning Works Operating Your Cleaner Before You Start Explanation of Controls Getting Ready Cleaning Items Optimizing Your Cleaner Application Hints Cleaning Solutions Troubleshooting Performance...
Safety Precautions Before using your Ultrasonic Cleaner, please read and thoroughly understand these safety precautions. Failure to follow them may result in serious personal injury or property damage. To avoid electrical shock: • Do unplug from power source before filling or emptying the tank. •...
Page 6
To prevent damage to the cleaner: • Do change your solution regularly. • Do not operate the cleaner dry. • Do not place parts or containers directly on the bottom of the cleaning tank; use a tray, wire hook or other device to suspend items. Failure to comply may cause transducer damage and will void your warranty.
PC620 General Information This PC-620 ultrasonic cleaner is available in two models; the unheated PC-620-1 and the heated PC-620-2. Each model is constructed using durable industrial style 44 kHz transducers. These provide increased cleaning power and ensure uniform cleaning activity throughout the bath.
Accessories For Your Cleaner A stainless steel cover and solid wall-perforated bottom parts basket are available for both models. Consult your distributor for price and delivery. Unpacking Your Cleaner Please check your cleaner and its carton carefully for any external or internal damage.
How Ultrasonic Cleaning Works Ultrasonic sound is sound transmitted at frequencies generally beyond the range of human hearing. In your ultrasonic cleaner, ultrasonic sound (sonics) is used for cleaning materials and parts. This is how it works: As the sound waves from the transducer radiate through the solution in the tank, they cause alternating high and low pressure...
Use only water-based solutions. • Do not ever use mineral acids. These could damage the tank. Failure to comply with these cautions will void your warranty. Operating Your PC-620-1 or PC-620-2 Cleaner Explanation of Controls Control...
Getting Ready Step Action Select your cleaning chemistry (check with your chemical supplier for solution effects on metals). Allowing for the volume of the parts you will be cleaning and cleaning chemistry, fill the tank with warm tap water to the operating level (one inch from the top).
Optimizing Your Cleaner Tanks Cleaning - check the tank for contamination whenever you change solution. If necessary, remove contaminants with a nonabrasive cloth and water. Emptying - always unplug the cleaner before emptying the tank. Dispose of the contents in an environmentally responsible way. Filling - always unplug the line cord before filling the tank.
Application Hints First time cleaning - first experiment with one piece, then proceed with the remainder. Solution level - Be sure to maintain solution level within one inch of the tank top. Surface activity will vary with liquid level. Load size - It is faster and more efficient to run several small loads rather than a few big loads.
Cleaning Solutions CAUTION Do not use alcohol, gasoline, bleach, mineral acids, solutions with a flash point, semi-aqueous or combustible liquids in ultrasonic tanks, or you will void the warranty. Only use non-flammable and water-based solutions. Solution Types Water-based solutions are either slightly acidic or alkaline. They include detergents, soaps and industrial cleaners designed to remove specific soils.
Page 16
Change the cleaning solution periodically. Cleaning solutions can become contaminated with soil particles which coat the tank bottom. This coating dampens the ultrasonic action and reduces cleaning efficiency. Certain solutions will cavitate better than others. Contact your local distributor for further information. Heat and cavitation: increase the chemical activity of cleaning solutions.
Troubleshooting If your cleaner does not operate satisfactorily, please check the tables below for possible causes before calling your authorized service center. WARNING High voltage inside - dangerous shock hazard. DO NOT attempt to disassemble or repair the cleaner. Problem Cause What to do Cleaner...
Performance Check your cleaner periodically to test the level of activity of the ultrasonic cavitation. Frequency of testing will depend on your use of the cleaner, however, we suggest running this test monthly. Glass Slide Test You will need the following equipment: •...
The ultrasonics will begin immediately to remove the lead from the slide. All lead should be removed within 10 seconds. If your cleaner passes this test, its ultrasonic cavitation is acceptable. NOTE: To ensure consistency from test to test, be sure to repeat test conditions - use the same solution concentration, liquid level, temperature, type of pencil, length of degassing, etc.
Need help?
Do you have a question about the PC-620-1 and is the answer not in the manual?
Questions and answers