Branson 1800 Operator's Manual page 40

Ultrasonic bath
Hide thumbs Also See for 1800:
Table of Contents

Advertisement

1.800.544.2843
Cleaning Items (Treating Samples) in Constant Sonics Mode
NOTE: To stop ultrasonics at any time, press the Sonics key (
Step
Action
1
Turn Main Power switch on.
2
Press the On/Standby key (
3
If necessary degas the liquid. See page 33 for instructions.
4
Change sonics mode:
• Press the Fn key until the Constant Sonics icon (
the Set Time icon (
• Press the
5
Set the tank temperature:
• Press the Fn key until the Set Temperature icon (
flashes.
• Press the
temperature you wish to reach.
• Press the HEAT key (
icon (
) appears.
NOTE: Units can be switched between °F or °C using the Fn
key. See Fn key description on page 29.
6
Set the ultrasonic power level:
• Press the Fn key until the Power Level icon (
• Press the
the
key to select low power ultrasonics.
7
Press the Sonics key (
8
Place the items into a basket, perforated tray, or beakers in a
positioning cover.
9
If using beakers or a solid tray, add cleaning solution to
beakers or tray to cover the items.
10
Slowly lower the tray or beakers into the tank. Do not allow
items to contact the tank bottom.
11
When items are clean, slowly remove them from the tank.
12
Rinse clean items with clean, warm water and dry, if
necessary.
35 Operating Your Ultrasonic Bath
www.calcert.com
15
10
20
5
25
0
30
) to turn on the unit.
) flash.
key to select Constant Sonics Mode.
/
keys to alter the setting to the tank
) once to activate heat. The Heat
key to select high power ultrasonics or press
) to activate ultrasonics.
).
) and
)
) flashes.
sales@calcert.com

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents