Operation
Command Syntax
•
Example:
Table 3.3 Command Parameters
Parameter
<current>, <Ihi>, <Ilo>
<time>
<voltage>, <Vlo>, <Vhi> The voltage in volts or millivolts. If no unit is
<mnemonics>
<state>
Floating Point Number <float> Variables sent with command parameters are
floating point numbers. Table 3.4 defines the structure of floating point numbers for
use with the software commands.
Table 3.4 Floating Point Numbers
Floating Number Definition
The floating point number has four significant fig-
ures. It can be of either sign, positive or negative.
A floating point number can have one decimal point.
Scientific Notation
Use E or e after the number for a base ten exponent.
An integer of either sign must follow an exponent.
34
Use commas between parameters in those commands with more than one
parameter, and between mnemonic parameters as in the MASK and UNMASK
commands. Only one comma is allowed and it may be preceded or followed by
any number of spaces.
MASK CV, OV, FOLD
Description
The current in amps or milliamps. If no unit is
given, the default unit is amps.
The time in seconds or milliseconds. If no unit
is given, the default unit is seconds.
given, the default unit is volts.
A combination of CV, CC, CV, OV, OT, SD,
FOLD, ERR, PON, REM, ACF, OPF, and
SNSP. See MASK and UNMASK commands
in the command reference for use of the ALL
and NONE parameters.
The state of a binary condition.
Example
1.234
-1.234
+1.234
0.123
1.2
123.4
123.0E-1
1.2E-1
10.00E+1
Operating Manual for GPIB for XHR/XFR Series Power Supply
Form
<float>
<float>A
<float>mA
<float>
<float>s
<float>ms
<float>
<float>V
<float>mV
See registers
on page 46.
<1/ON,
0/OFF>