Care and maintenance
Check LED 3 ("Level indicator")
Switch off the main switch
Take out module 2
Unplug centrifuge connector SV2 on the main board (1)
On the main board, cover the lower diode of the light barrier (2)
Switch on the main switch (3)
LED 3 lights up yellow:
Buzzer sounds (switchable via RESET)
3
Switch off the main switch
On the main board, cover both diodes of the light barrier (2)
Switch on the main switch (3)
LED 3 lights up yellow:
Buzzer sounds (cannot be switched off)
With module 2 inserted, suction is now not possible
Switch off the main switch
Reinsert centrifuge connector SV2
Reinsert module 2
4
Carry out a normal operation test
7.2.5.
5-year service kit
According to the German Waste Water Ordinance, Annex 50, amalgam separators must be inspected for proper condition at intervals
i
of no longer than 5 years in accordance with state law!
The 5-year inspection must be entered in the device document!
1
Carry out a 1-year inspection
Check the correct installation and connection of the amalgam separator in accordance with the
2
installation requirements
Carry out disinfection and cleaning of the suction hoses and spittoon
3
4
Carry out a visual inspection of the centrifuge chamber
5
Carry out a normal operation test
7.2.5.1. Visual check of the centrifuge
Tilt the locking bracket on module 2 upwards and take out module 2.
1
Remove locking clip on centrifuge support.
2
26
2
1
3
see 7.2.3. Normal operation test
see 7.2.4. 1-year service kit
see 5.2. Installation requirements
see 7.1.1. Daily cleaning with GREEN&CLEAN
M2
see 7.2.5.1. Visual check of the centrifuge
see 7.2.3. Normal operation test