CPI Millenia Series Operator's Manual page 40

X-ray generator

Advertisement

6
Error Codes and Messages
ERROR CODE
E016
E017
E018
E019
E020
E021
E022
E023
E024
E025
E026
E028
E029
E030
E031
E032
E033
E034
E035
Page 40
MESSAGE
TOMO BUT ERROR
NOT CALIBRATED
PREP TIMEOUT
ANODE HEAT LIMIT
THERMAL INT #1
THERMAL INT #2
DOOR INTERLOCK
COLLIMATOR ERR
CASSETTE ERROR
II SAFETY INT
SPARE INT
PREP SW CLOSED
X-RAY SW CLOSED
FLUORO SW CLOSED
REMOTE COMM ERR
CONSOLE COMM ERR
GEN BATTERY LOW
+12VDC ERROR
-12VDC ERROR
Rev. H
Millenia / Indico 100 Series Operator Manual
PROBLEM
Tomo exposure exceeded
back up time.
Generator has been in prep
state too long.
Selected parameters will
cause X-ray tube to exceed
its programmed anode heat
limit.
X-ray tube # 1 too hot and
its thermal switch has
opened.
X-ray tube # 2 too hot and
its thermal switch has
opened.
Door is open.
Collimator is not ready.
Cassette is not ready.
II safety is not ready.
Spare input is not ready.
Communication error
detected with remote fluoro
control unit.
Generator has detected
error in communication to
console.
CPI Canada Inc
ACTION
1. Check exposure
technique settings.
2. Increase tomo
back up time if
necessary.
Consult your service
representative.
Reduce length of
time in prep state.
Reduce parameters
or wait for tube to
cool.
Wait for X-ray tube #
1 to cool.
Wait for X-ray tube #
2 to cool.
Close door.
Check collimator.
Check cassette.
Check II safety.
Consult your service
representative.
Consult your service
representative.
Consult your service
representative.
Consult your service
representative.
Turn power off and
then on to Reset
Generator.
Consult your service
representative.
Turn power off and
then on to reset
generator.
Consult your service
representative.
Consult your service
representative.
Consult your service
representative.
Consult your service
representative.
740811-00

Advertisement

loading

This manual is also suitable for:

Indico 100 series