Console Controls
CPI Canada Inc
ANATOMICAL PROGRAMMING CONTROLS AND DISPLAY
THE ABOVE IS A TYPICAL DISPLAY. YOUR UNIT MAY DIFFER IN DETAILS
1. Anatomical section/anatomical view selection buttons
2. Anode heat indicator (%) readout
3. Time readout - 24 hour mode (RAD only)
Displays mag mode in fluoro
4. Status and operator message display area
5. ms/mAs readout
6. Dose readout - fluoro only
7. Back and next selection buttons
8. Reset button.
Press the button next to the anatomical section required. The APR display will change to show
anatomical views within that section. Press the button for the required function. The
radiography display will show the programmed X-ray technique factors.
Press the function button again to toggle between normal, large and small patient sizes. The
default is normal.
You may override the programmed techniques by selecting + or - for the kV, mA or AEC
density. This will not change the programmed parameters. To make permanent changes to
the APR, see CHAPTER 5: PROGRAMMING THE CONSOLE.
%HU indicates the tube anode heat for the selected X-ray tube. An anode heat warning
message will be displayed at an installer-programmable level, typically 80 %. Exposures that
will exceed a value of typically 90 % will be inhibited (this is also installer programmable).