X-ray generator

Advertisement

CPI Canada Inc
Error Messages
These messages indicate that an error has occurred.
ERROR CODE
E001
E003
E004
E005
E006
E007
E008
E009
E011
E012
E013
E014
E015
Millenia / Indico 100 Series Operator Manual
MESSAGE
GEN EPROM ERR
GEN NVRAM ERR
GEN RTC ERROR
PS CONTACT ERR
ROTOR FAULT
FILAMENT FAULT
MA FAULT
PS NOT READY
HIGH MA FAULT
LOW MA FAULT
MANUAL TERMIN
AEC BUT ERROR
AEC BU MAS ERR
PROBLEM
Generator CPU real time
clock is not functioning.
Power supply is not ready to
make an exposure.
Operator released exposure
switch during exposure.
AEC exposure exceeded
allowed back up time.
AEC exposure exceeded
allowed back up mAs.
740811-00
Rev. H
Error Codes and Messages
ACTION
Consult your service
representative.
Consult your service
representative.
Reset time and date.
Consult your service
representative.
Consult your service
representative.
Power unit off and
retry rotor start.
Consult your service
representative.
Consult your service
representative.
Consult your service
representative.
Retry exposure.
Consult your service
representative.
Consult your service
representative.
Consult your service
representative.
Re-take exposure if
necessary.
Consult your service
representative.
Check exposure
technique settings.
Consult your service
representative.
Check exposure
technique settings.
Consult your service
representative.