The current time is displayed in 24 hour format in the upper right corner of the APR window.
The time display is replaced by a mag mode indicator when in fluoro operation.
The lower center part of the APR display shows status, warning, and error messages.
The significance of the ms/mAs readout varies depending on mode selected:
• In AEC mode, with fixed backup time selected, the AEC backup time is displayed.
• In AEC mode, with either variable ms or variable mAs backup time, the backup mAs or ms
value is displayed.
• In mAs mode, the time of the exposure will be displayed.
• In mA/ms mode, the calculated mAs of the exposure will be displayed (based on the mA
and ms selected).
Press the BACK or NEXT buttons to scroll forward or back through the display. An appropriate
message, i.e. NEXT/BACK or equivalent will be displayed when these buttons are active.
Press the RESET button to return to the main menu, and clear status and error messages.