Technical Information

Window Description

Table 7
Window meanings
Win #
Read
Write
0
Always
Serial
1
Always
Serial
8
Always
Always
Stop &
100
Always
Serial
101
Always
Always
102
Always
Always
103
Always
Always
104
Always
Always
105
Always
Always
108
Always
Always
109
Always
Always
110
Always
Always
111
Always
Always
117
Always
Always
120
Always
Always
121
Always
Always
122
Always
Always
Agilent TwisTorr 305-ICQ & 305-IC SF User Manual
Type
Name
L
Start/Stop
L
Low Speed
A
Serial/Remote
L
Soft Start
N
Setpoint Type
N
Setpoint Value
N
Setpoint Delay
L
Setpoint Logic
N
Setpoint histeresis
N
Baud Rate
L
Pump Life Reset
L
Interlock Type
Analog Output
N
Type
Low Speed Adjust
N
Frequency
N
Target Frequency
Maximum
N
frequency
L
Set vent valve
Meaning
Start/Stop command
Low Speed command
Configures controller
Enable/Disable Soft Start option
It defines set point type:
0 = Freqency
1 = Power
2 = Time
3 = Normal
It defines the threshold value for Set Point
It defines the setpoint mask time from start
It defines the functionally logic. "High level active" indicates an
output ON when the signal is above the threshold and "Low level
active" is the opposite.
It defines the percentage value of Set Point Hysteresis
It defines the serial communication baud rate:
0 = 600 bit/s
1 = 1200 bit/s
2 = 2400 bit/s
3 = 4800 bit/s
4 = 9600 bit/s
5 = 19200 bit/s
6 = 38400 bit/s
Pump Life Reset
It defines the interlock type. The interlock can be checked at start
or continuously
It defines the analog output type:
0 = Frequency
1 = Power
2 = Body Temperature
It defines the Low Speed frequency value
It defines the target excitation frequency
It defines the maximum frequency settable by WIN120
Set vent valve
Default
(Factory
Set)
Stop
Off
Remote
Off
Frequency
720
0
Direct
2
9600 bit/s
---
Continuous
Frequency
700Hz
1010Hz
1010Hz
Off
257