TwisTorr 305-ICQ & 305-IC SF Mounting
Failure to comply with these installation instructions could result in the
WARNING
pump detaching from the system in the event of a rotor failure, which
could cause property damage or serious injury or death.
The TwisTorr 305-ICQ & 305-IC SF can be installed in any position.
Fix the TwisTorr 305-ICQ & 305-IC SF in a stable position, mounting the inlet flange
of the turbopump to the system counter-flange, with a connection capable of
withstanding a torque of 1000 Nm around its axis.
The turbopump with ConFlat and ISO K inlet flange must be fixed to the vacuum
chamber by means of the appropriate Agilent hardware.
For further details see the "Installation of ConFlat and ISO K" section.
The TwisTorr 305-ICQ & 305-IC SF cannot be fixed by means of its base or
CAUTION
through other flanges that are not inlet.
Agilent TwisTorr 305-ICQ & 305-IC SF User Manual