Purge Valve Installation - Agilent Technologies TwisTorr 305-ICQ User Manual

Table of Contents

Advertisement

Technical Information

Purge Valve Installation

A gas purge valve is available to protect the pump bearings against particulate and
corrosive gases that could move into the pump. To install the gas purge valve it is
necessary to switch off the pump and unscrew the purge port cover as shown in
the following figure.
Figure 77
and then screw the gas purge valve (with a torque of 2.5 Nm) as shown in the
following figure.
Figure 78
Agilent TwisTorr 305-ICQ & 305-IC SF User Manual
241

Advertisement

Table of Contents
loading

Table of Contents