Appendix 1 - Helpful Tips - Factair F6100 Operating Instructions Manual

Safe-air monitor
Table of Contents

Advertisement

 Never connect to a non-regulated supply from an HP cylinder or compressor.
 Maximum inlet pressure to the Safe-Air Monitor is 10 bar. If exposed to an overpressure
the instrument will display an overpressure warning necessitating its return to the
manufacturer for checking and resetting.
 The off button is located within the configuration menu and is deliberately not available on
the home screen or sub menu to prevent accidental powering down the unit.
 The following warning will be displayed if there is insufficient pressure being supplied for
the tube test
 For any additional advice and information please contact the manufacturer (see details on manual
cover).
 1 minute prior to the test results being saved onto the SD card a blinking icon will be displayed on
the home screen. This icon will continue to blink if the SD card is not in the instrument and the
F6100 has been unable to save the results.
HELPFUL TIPS
To turn the instrument off press the
button in the top right hand corner of the
screen.
The instrument requires a minimum of 2
bar to operate, if this warning is
displayed there may also be insufficient
pressure for the electronic sensors to
operate correctly.
18
APPENDIX 1