ResMed AirMini User Manual page 32

Hide thumbs Also See for AirMini:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of the
specified air tubing:
Pressure
cm H
O (hPa)
2
4
8
12
16
20
Measurement system uncertainties
In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the
manufacturer's test equipment is:
For measures of flow
For measures of static pressure
For measures of dynamic pressure
For measures of volume (< 100 mL)
For measures of volume (≥ 100 mL)
For measures of time
Note: ISO 80601-2-70:2015 stated accuracies and test results provided in this manual for
these items already include the relevant measurement uncertainty from the table above.
Sound
Declared dual-number noise emission values in accordance with ISO 4871:1996
Pressure level measured according to
ISO 80601-2-70:2015 (CPAP mode)
Power level measured according to
ISO 80601-2-70:2015 (CPAP mode)
The A-weighted sound pressure level measured according to ISO 17510:2015:
Air tubing connected to AirMini F20 connector
with mask
30
AirMini machine and AirMini air tubing
L/min (including uncertainty)
119
116
112
108
105
± 1.5 L/min or ± 2.7% of reading (whichever is
greater)
± 0.15 cm H
± 0.27 cm H
± 5 mL or 6% of reading (whichever is greater)
± 20 mL or 3% of reading (whichever is greater)
± 10 ms
O (hPa)
2
O (hPa)
2
29 dBA with uncertainty of 2 dBA
37 dBA with uncertainty of 2 dBA
19 dBA with uncertainty of 3 dBA

Advertisement

Table of Contents
loading

Table of Contents