7 series Blood Pressure Monitor
Model BP761N
Instruction Manual
Spanish instruction manual can be found on
OmronHealthcare.com.
2298830-6D
Product includes:
Battery Set
ComFit™ Cuff
Monitor
GETTING
STARTED
Blood Pressure Monitor
BP761N
2298832-2C
Quick Start Guide
Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 7 Series and is the answer not in the manual?

Questions and answers

MrsF
April 30, 2025

Why do I get E5 error “clothing is interfering” sometimes when my arm is bare?

1 comments:
Mr. Anderson
May 16, 2025

If the E5 error indicates "clothing is interfering" but the arm is bare, it may be a misinterpretation of the actual cause. E5 typically means the pulse rate is not detected correctly. This can happen due to improper cuff placement, excessive movement, or poor contact with the skin, even if no clothing is present. Ensure the cuff is wrapped correctly and snugly around the bare arm, and stay still during measurement.

This answer is automatically generated

Janet
July 21, 2025

How many times should I take my B.P. to get the most accurate reading? I have heard that it is necessary to take it twice (after “resting”a few minutes between takes. Have also heard that 3 times is even more accurate. I’m sorry, but I could not find the answer to that question in the manual.

Sarah lee Oliver
April 5, 2025

I can't find the adjusting screw

Summary of Contents for Omron 7 Series

  • Page 1 Product includes: Battery Set ComFit™ Cuff Monitor Manufactured for: OMRON HEALTHCARE Co., Ltd. 7 series Blood Pressure Monitor GETTING 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN STARTED Blood Pressure Monitor Distributed by: Model BP761N BP761N OMRON HEALTHCARE, INC. 2298832-2C Instruction Manual Quick Start Guide 1925 West Field Court Lake Forest, IL 60045 U.S.A.
  • Page 2: D A Important Safety Information

    Important Safety Information Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. The monitor is not intended to be a diagnostic device.
  • Page 3 Important Safety Information Caution: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. (General Usage) Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous. Consult your physician before using the device for any of the following conditions: •...
  • Page 4 Wipe the dust off from the power plug. Unplug monitor when not in use. Disconnect the power plug before cleaning. Use only an OMRON AC adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device. (Battery Usage) Do not insert the batteries with their polarities incorrectly aligned.
  • Page 5: Know Your Device

    1. Know Your Device Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. Monitor Display Air jack Memory symbol Blood pressure color indicator USER ID symbol Connection button Average value symbol...
  • Page 6: Display Symbols

    1. Know Your Device 1.1 Display Symbols Irregular Heartbeat Symbol ( Normal Heartbeat When the monitor detects an irregular rhythm two or more times during the Pulse measurement, the irregular heartbeat symbol will appear on the display with Blood pressure the measurement values.
  • Page 7: Before Taking A Measurement

    1. Know Your Device Blood Pressure Color Indicator If your Systolic Blood Pressure is 135 mmHg or above and/or the Diastolic Blood Pressure is 85 mmHg or above, the blood pressure color indicator will light in “orange” when the measurement result is displayed. If the measurements are within the standard range, the blood pressure color indicator will light in “green”.
  • Page 8: Preparation

    2. Preparation 2.1 Battery Installation Remove the battery cover. Insert 4 “AA” batteries as indicated in the battery compartment. Replace the battery cover. Notes: • When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time.
  • Page 9: Setting The Date And Time

    2. Preparation Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. 2.2 Setting the Date and Time The date and time will be automatically set when paired with your smartphone. To set manually, follow the instructions below.
  • Page 10 2. Preparation Push or button to change the year. Push button to confirm the year and then the month flashes. Repeat the same steps to change the month, day, hour, and minute. Year Month Hour Minute Change Confirm Change Confirm Change Confirm Change Confirm...
  • Page 11: Using The Device

    • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your...
  • Page 12: How To Sit Correctly

    3. Using the Device 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, at a comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. •...
  • Page 13: Taking A Measurement

    3. Using the Device 3.3 Taking a Measurement Notes: • To stop a measurement, press the START/STOP button once to deflate the arm cuff. • Remain still and quiet while taking a measurement. The monitor is designed to take measurements and store the measurement values in the memory for 2 people using USER ID 1 and USER ID 2.
  • Page 14 3. Using the Device Remove the arm cuff. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes. Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.
  • Page 15 3. Using the Device DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. This monitor is not intended to be a diagnostic device. Always consult your physician.
  • Page 16: Using The Memory Function

    3. Using the Device 3.4 Using the Memory Function The monitor automatically stores the results up to 60 readings for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes.
  • Page 17 3. Using the Device Press the button to view the values stored in the memory. : To view the older values : To view the more recent values To View the Average Value Select your USER ID (1 or 2). Press and hold the button for more than 3 seconds.
  • Page 18 To Pair This Device With a Smartphone Review the list of compatible smart devices at: OmronHealthcare.com To begin, download and install the free "Omron connect US/CAN" App onto your smartphone. Open the App on your smartphone and follow set-up and pairing instructions.
  • Page 19 3. Using the Device Open the "Omron connect US/CAN" App on your smartphone and follow the device pairing instructions shown on your smartphone, while the connection symbol is flashing on the device’s display. Confirm if the device is connected successfully.
  • Page 20 Clear your reading screen by pressing the "START/STOP" button on your monitor. Next, tap the "transfer" icon in the Omron connect US/CAN app to start the transfer. * If the data transfer is not starting, refer to “Connection failure. Data cannot be sent.”...
  • Page 21 3. Using the Device TO DELETE ALL RESULTS AND SETTINGS To delete all the information stored in your monitor, follow the instructions below. Make sure that the monitor’s power has been off. While holding the button down, press the START/STOP button for more than 5 seconds.
  • Page 22 3. Using the Device To Delete All the Values Stored in Memory The values stored in the memory are deleted by USER ID. Select your USER ID (1 or 2). Press the Memory button, while the memory symbol ( ) appears.
  • Page 23: Turning "Off" The Bluetooth

    3. Using the Device 3.5 Turning “OFF” the Bluetooth ® Turn off the Bluetooth ® of the monitor in the following areas where use of wireless equipment is prohibited. • On aircraft • In hospitals •While abroad Make sure that the power has been off before starting the instructions below. Press and hold the connection button for more than 10 seconds.
  • Page 24 3. Using the Device ® To turn “ON” the Bluetooth Make sure that the power has been off before starting the instructions below. Press and hold the connection button for more than 10 seconds. Release the button when “ON” appears on the display. Press the START/STOP button to turn the monitor off.
  • Page 25: Error Messages And Troubleshooting

    4. Error Messages and Troubleshooting 4.1 Error Messages Display Cause Solution Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Irregular heartbeats are detected. Repeat the steps in section 3.3. If this error continues to appear, contact your physician. Carefully read and repeat the steps in section Movement during measurement.
  • Page 26 4. Error Messages and Troubleshooting Display Cause Solution Insert the air plug securely. Air plug disconnected. Refer to section 3.1. Apply the arm cuff tighter. Arm cuff is applied too loosely. Refer to section 3.1. Replace the arm cuff with a new one. Air is leaking from the arm cuff.
  • Page 27: Troubleshooting

    Connection failure. Refer to "To Pair This Device With a Smartphone" in Data cannot be sent. section 3.4 or visit the "Support" section in the Omron connect US/CAN app for further help in pairing your smart device. The blood pressure monitor is not ready, follow the step “1.
  • Page 28: Maintenance And Storage

    5. Maintenance and Storage 5.1 Maintenance To protect your device from damage, please observe the following: • Store the device and the components in a clean, safe location. • Do not use any abrasive or volatile cleaners. • Do not wash the device and any components or immerse them in water. •...
  • Page 29: Storage

    5. Maintenance and Storage 5.2 Storage Unplug the air plug from the air jack. Gently fold the air tube into the arm cuff. Note: Do not bend or crease the air tube excessively. Do not store the device in the following situations: •...
  • Page 30: Optional Medical Accessories

    5. Maintenance and Storage 5.3 Optional Medical Accessories Arm Cuff AC Adapter Arm circumference 9” - 17” (22 - 42 cm) HEM-ADPTW5 ComFit Cuff (CFX-WR17) (Model: HEM-FL31)
  • Page 31 5. Maintenance and Storage Using the Optional AC Adapter Insert the AC adapter plug into the AC adapter jack on the rear side of the monitor. Plug the AC adapter into an electrical outlet. To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the monitor.
  • Page 32: Specifications

    6. Specifications Model HEM-7320T-Z BP761N Display LCD digital display Measurement range Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min. Accuracy Pressure: ±3 mmHg or 2% of reading Pulse: ± 5% of display reading Inflation Fuzzy-logic controlled by electric pump Deflation Automatic pressure release valve Measurement method...
  • Page 33: Fcc/Ic Statement And Trademarks

    7. FCC/IC Statement and Trademarks FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
  • Page 34 The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by OMRON HEALTHCARE Co.,Ltd. is under license. Other trademarks and trade names are those of their respective owners. Apple, the Apple logo, iPad, and iPhone are trademarks of Apple Inc., registered in the U.S. and other countries.
  • Page 35: Limited Warranty

    Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend insuring the product with return receipt requested. THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
  • Page 36: Guidance And Manufacturer's Declaration

    Nevertheless, special precautions need to be observed: • The use of accessories and cables other than those specified by OMRON, with the exception of cables sold by OMRON as replacement parts for internal components, may result in increased emission or decreased immunity of the device.
  • Page 37 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP761N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP761N including AC-adapter should assure that it is used in such environment.
  • Page 38 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP761N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP761N including AC-adapter should assure that it is used in such environment.
  • Page 39 OMRON BP761N including AC-adapter OMRON BP761N including AC-adapter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this OMRON BP761N including AC-adapter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the OMRON BP761N including AC-adapter as recommended below, according to the maximum output power of the communications equipment.
  • Page 42: Table Of Contents

    Contents Thank you for purchasing the OMRON ® BP761N Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.