4. Select a Stimulation Set and increase current to the patient preferred level, then stop
stimulation.
5. Repeat Step 4 for each remaining Stimulation Set in the Program until all have been
set to the patient preferred level (Figure 7.21).
6. Set the Stimulation Sets to run in Grouped mode (
increase (all Stimulation Sets) to the patient preferred level.
7. Enable Feedback Control as per Section 7.11.
Note: Feedback Control will only operate based on the settings in Stimulation Set A
(it will vary the current so that the ratio of current between Stimulation Sets remains
constant).
Figure 7.21: A Program with 4 Stimulation Sets configured at desired levels. Note that the
electrode settings from all Stimulation Sets are visible in the background, while the active
7.13.2 Enable/Disable a Stimulation Set in a Program
To enable/disable a Stimulation Set in a Program:
1. Select the Stimulation Set you want to enable/disable.
2. Press
to enable the Stimulation Set or
Stimulation Set will have a red diagonal line across it, e.g.
3. Start stimulation again when the desired changes have been made.
Note: The Stimulation Sets that are still enabled will continue to increase at the
previous ratio (Figure 7.22).
Note: If a Stimulation Set is disabled (
to be reconfigured (refer to Section 7.13.3).
CLIN-UMAN-002424
Document Reference:
Evoke™ Clinical Manual
set is displayed as normal.
) its ratio will be set to 0 and it will need
Revision: 12.00
), start stimulation and
to disable it. A disabled
.
Page 25 of 56