Table 11: Error Codes - Nellcor OxiMax N-550B Service Manual

Table of Contents

Advertisement

Troubleshooting
50
displays a red "EEE" in the left numeric display area
displays a red error code in the left numeric display
Note: Cycling the power clears the displayed error code
Error
Meaning
Code
1
SpO
front end RAM error
2
2
SpO
front end ROM/code integrity error.
2
3
SpO
front end reported a bad CRC
2
4
SpO
front end reported FSP message not allowed
2
5
SpO
front end reported illegal value sent in FSP messageSpO
2
end
6
SpO
front end reports calibration (offset) failure
2
9
SpO
front end reported syntax error in FSP message
2
10
-
Over
current limit in SpO
11
SpO
front end reports incorrect system voltage
2
12
SpO
front end reports other hardware problem
2
14
SpO
front end reports communication channel overflow
2
16
SpO
front end reports watch dog time out
2
17
SpO
front end reports that sensor appears defective
2
18
SpO
front end reports internal register appears modified from
2
expected value
19
SpO
front end reports signal out
2
48
SpO
front end reports spurious interrupt
2
49
SpO
front end reports internal buffer overflow
2
50
SpO
front end reports intermittent error
2
51
SpO
front end reports digital communications error
2
52
SpO
front end reports warmer error
2
53
Front end data not received
256
SpO2 back end reports beginning of packet missing
257
SpO2 back end reports packet start ID (SID) missing

Table 11: Error Codes

front end has tripped
2
-
-
of
range
front
2

Advertisement

Table of Contents
loading