en
Description
2.5
Approvals
The closed-circuit breathing apparatus is approved in
accordance with Gost R 53259-2009 with the following
facepieces:
Dräger FPS 7000 RP Series
Panorama Nova RP Series
It conforms to the following regulations:
Gost R 53259-2009
(EU) 2016/425
TR CU 019/2011
Declaration of conformity:
see www.draeger.com/product-certificates
The PSS BG 4 RP is approved in line with Poshtest.
Depending on which apparatus Bodyguard is installed in,
the Bodyguard will fulfil the explosion protection requirements
of the following standards:
Bodyguard +
Conforms to
apparatus
regulation
PSS BG 4 RP
TR CU 019/2011
44
2.6
The following information can be shown on the Bodyguard
display:
Symbol
Bodyguard
marking
T
= -40 °C to
a
60 °C
PO ExiaI X
O ExiaIICT4 X
900
The LEDs have the following meanings:
LED
Green LED
flashing
Red LEDs
flashing
Explanation of symbols
Meaning
Test completed successfully.
An error has occurred.
The error is indicated by the letter in front of
the X.
Further information can be obtained from
Dräger.
The numbers to the left of this symbol give
the remaining period of use in minutes.
The leak test is currently running.
Infrared connection between Bodyguard
and IR interface.
The function key has been inserted.
The motion sensor is disabled.
The function key has been removed.
The motion sensor is activated.
Battery symbol
first battery warning
second battery warning
Open cylinder valve!
Close cylinder valve!
Country code (information on which cylinder
threaded connection is programmed in the
Bodyguard):
900: G3/4"
Meaning
The Bodyguard is in normal operating
mode.
The Bodyguard is issuing an alarm or an
error has occurred during the self-test.