...10 PROGRAMMING

10.3 Set Up Outputs Page

Note. Access to the programming pages is possible only with Range 1 selected.
SET UP OUTPUTS
Alarm 1 Set Point
Alarm 2 Set Point
RTX Type
0-10
0-20
4-20
RTX Range
85%
xxx.x
Test Retrans (%)
Alter Sec. Code
Alarm 1 Set Point
The set point band is defined as the actual value of the set point plus or minus the
hysteresis value. The hysteresis value is fixed at 0.2% H
input value is above or below the set point band. If the input moves within the set point
band, the last alarm action is maintained.
The
can be set to any value within the input range being displayed. The
Alarm 1 Set Point
decimal point position is set automatically. The alarm LEDs are illuminated in the alarm
condition.
Alarm 2 Set Point
See
.
Alarm 1 Set Point
Advance to next parameter
Retransmission Output Type
Select retransmission output range required (4 to 20mA, 0 to 20mA or 0 to 10mA).
Advance to next parameter
RTX Range
The retransmission signal can be selected to be ranged 85 to 100% H
100% H
in Air. This option is only available with the range switch in position 1.
2
Advance to next parameter
Test Retransmission Output
The instrument transmits a test signal of 0, 25, 50, 75 or 100% of the retransmission
range. The % test signal selected is shown on the upper line of the display.
– for the range 4 to 20mA and 50% retransmission test signal, 12mA is
Example
transmitted.
Select the required retransmission test signal.
Advance to next parameter
Alter Security Code
Set the security code to a value between 00000 and 19999.
This value will then have to be entered to regain access to the secure parameters.
Advance to
Electrical Calibration Page
in Air. Alarm action occurs if the
2
.
in Air or 80 to
2
29