EM Pointer
18099-02A
1
ENGLISH
General Information
Ensure patient safety by following these instructions for using
EM Pointer:
EM Pointer is used to:
1.
Verify registration accuracy
Navigate
Calibrate additional instruments
2.
EM Pointer is precalibrated and can be used without
further calibration.
3.
EM Pointer is not intended for use in an MR
environment. Do not expose the device to a highly
magnetic field, such as an MRI scanner, as it may
become magnetized. Tracking with a magnetized device
may result in incorrect navigation and possible injury to
the patient and/or medical personnel.
The EM Pointer surface must be sterilized prior to use.
4.
EM Pointer Components (fig. 1):
No.
Component
Function/Usage
Calibration of additional instruments
Calibration
Point
using an EM Instrument Reference
Control
Activates various functions, as
Button
described in the Software User Guide
Cable
Data transfer
Connects to the EM Tracking Unit
Plug
using an instrument port on the EM
Base Station
How to switch on EM Pointer
Attach the EM Pointer plug to an instrument port on the EM
Base Station.
After the corresponding LED on the EM Base Station turns
green, EM Pointer is ready for use.
NOTE: It may take a few moments for the LED to turn green.
How to switch off EM Pointer
Pull the EM Pointer plug from the EM Base Station by slightly
pulling at the chequered area of the plug.
Point Acquisition Procedure
Place the tip of the EM Pointer on the point of interest and
press the control button to acquire the point. When EM
Pointer acquires a point, the exact pointer tip position is
transmitted to the EM Tracking Unit.
Warning
Use EM Pointer only as described and only in combination
with devices that are specified in the System User Guide.
Warning
Only use EM Pointer after verifying its accuracy. The
verification point on a sterile EM Instrument Reference can
be used for verification. Do not use the verification point on
the EM Patient Reference, as the EM Patient Reference
may not be sterile.
Warning
If EM Pointer verification is unsuccessful, do not use it for
surgery and contact Brainlab support.
Brainlab AG
Customer Support:
Kapellenstraße 12
support@brainlab.com
85622 Feldkirchen
North/ South America: +1 800 597 5911
Germany
Europe&Rest of the world: +49 89 99 15 68 44
2
Warning
Do not use EM Pointer on the patient's eyes.
Warning
Be aware that pressing the control button may cause the
pointer tip to change position.
Warning
Hold the pointer steady to avoid movement when pressing
control button.
Warning
Do not run the cable over the patient's throat as this could
lead to patient injury.
Warning
Handle the EM Pointer with care, placing the cable in a way
that prevents tangling or tripping over it.
Warning
Do not wind the cable around the EM Field Generator as
this can cause the device to malfunction.
Warning
Do not bend or kink the cable and do not use EM Pointer if
the cable is damaged. Using the device with a damaged
cable could result in personal injury.
Warning
Do not drape EM Pointer, otherwise navigation might not
work or accuracy could be affected.
Warning
EM Pointer is a highly accurate and sensitive medical
instrument. To maintain calibration precision, it must be
handled with extreme care. If it is dropped or otherwise
mishandled (even when inside the sterilization tray), verify
the calibration accuracy as described in the Software User
Guide or return it to Brainlab for testing, replacement or
repair.
Warning
EM Pointer was not designed or tested for use during or
following cardiac defibrillation. Cardiac defibrillation may
affect the system accuracy or may damage the system.
Caution
Do not attempt to modify EM Pointer.
Caution
Do not expose EM Pointer to X-rays, as this could damage
the device.
Caution
Do not expose EM Pointer to excessive UV light like direct
sunlight, as this may affect the plastic housing.
Caution
Always store EM Pointer in the sterilization tray provided.
Caution
Only dispose of electrical and electronic equipment in
accordance with statutory regulations.
For information regarding the WEEE (Waste Electrical and
Electronic Equipment) directive, visit:
http://www.brainlab.com/weee
a
page 1/5
For further information please see Brainlab's user manual
or consult Brainlab customer support.
Reprocessing
1. Remove coarse impurities from EM Pointer directly after
use (within 2 hours) using running water (20-30°C) or a
cleaning-disinfectant solution.
2. Using a brush, apply detergent solution to all surfaces,
ensuring that all parts are cleaned.
• To manually remove impurities, use a soft brush or a
clean soft tissue
• Do not use metal brushes or steel wool
Cut off the next number from the sterilization counter
tag (fig. 2).
Warning
Dispose of EM Pointer when it has been reprocessed
20 times (i.e., all numbers are cut off).
3. Automatic cleaning and disinfection
4. Steam sterilization:
• Prevac procedure, 132°C (270°F), 5 minutes, drying
time at least 10 minutes
• Use EM Pointer Sterilization Tray.
The sterilization temperature must not drop below the specified
minimum temperature.
The stated sterilization time is the minimum validated
sterilization time required to achieve a 10-6 sterility assurance
level (SAL), and requires complete attention to the instructions
on pretreatment, cleaning and disinfection, as well as post-
drying and check of the instruments.
Warning
Any deviation from the validated parameters could result in
patient infection or injury.
NOTE: For further information, see the Brainlab Cleaning,
Disinfection and Sterilization Guide.
BL-IL-18099-02A Rev000
*18099-02A*

Advertisement

loading

Summary of Contents for Brainlab EM Pointer

  • Page 1 • Use EM Pointer Sterilization Tray. EM Pointer Components (fig. 1): Warning The sterilization temperature must not drop below the specified Do not bend or kink the cable and do not use EM Pointer if Component Function/Usage minimum temperature. the cable is damaged. Using the device with a damaged...
  • Page 2 Desinfektion sowie Trocknen und Überprüfen der Instrumente wird die exakte Position der Pointer-Spitze an die EM- beschrieben überprüft werden oder das Gerät zum Testen, sind genauestens einzuhalten. Tracking-Einheit übertragen. Austausch oder zur Reparatur an Brainlab zurückgeschickt werden. Warnung Warnung Jegliche Abweichung von den validierten Parametern kann...
  • Page 3 Posizionare la punta del puntatore EM sul punto di interesse e premere il pulsante di controllo per acquisire il punto. Quando il Brainlab per il collaudo, la sostituzione o la riparazione. NOTA: per ulteriori informazioni, consultare il Manuale di puntatore EM acquisisce un punto, la posizione esatta della pulizia, disinfezione e sterilizzazione Brainlab.
  • Page 4 Para obtener más información, consulte el manual de Cuando pulse el botón del control, mantenga el puntero antes de utilizarla. usuario de Brainlab o póngase en contacto con el Servicio inmóvil para evitar que se mueva. Técnico de Brainlab. Componentes del Puntero EM (fig. 1):...
  • Page 5 Guide d’utilisation du Avertissement Mettre le pointeur EM hors tension logiciel ou le renvoyer à Brainlab pour un test, un Tout changement par rapport aux paramètres validés peut Débranchez la fiche du pointeur EM de la station d’accueil remplacement ou une réparation.

This manual is also suitable for:

18099-02a