Self Cleaning
This is a feature for Belmont Medical Technologies certified technicians only. It is
password-protected.
This feature performs a thermal disinfection of the water tank and internal tubing.
The thermal disinfection of Allon
circulating water of the system, thus allowing the heat to disinfect the internal
water pathways of the system, including the water tank.
Thermal disinfection is performed at every Periodic Maintenance.
CAUTION!
NOTE:
Thermal Disinfection Process
Required Equipment
– Bypass tube PN #200-00181 or PN #200-00096
– Up to 8 liters of .22 µm filtered or sterile water
To perform Thermal Disinfection:
NOTE:
1. In the main menu, select Services.
2. Touch Self Cleaning and then OK.
3. The process is password protected. Insert password.
DDT-063-000 Rev C
®
is an integrated feature, which heats the
– Use only sterile or 0.22 µ filtered water.
– Do NOT use bleach or any other cleaning and disinfectant agent
for the internal circulation except sodium dichloroisocyanurate
(NaDCC). These agents may harm the system and result in
damage to the system.
– Always drain the water after the thermal disinfection process.
Self Cleaning is password protected and should only be used by
authorized Belmont Medical Technologies personnel.
Make sure that water tank is full, and that the bypass tube is
connected.
Belmont Medical Technologies
Allon
User Manual
®