Section 4 - Specifications Page; Optical Power Meter Specifications - Textron Tempo 525N-60 Operating Manual

Single-mode optical line test set
Table of Contents

Advertisement

Specifications

Optical Power Meter Specifications

Power Measurement Range in dBm
Wavelength > 850 - 1700nm
Wavelength Range
Calibration Points
Absolute Accuracy
1310nm with -10dBm input power
Linearity @ 1310nm
Linearity Accuracy
+10 to -3dBm
-3 to -55dBm
-55 to -65dBm
Optical stability
OPM Channel only
Settling time, auto-range
Optical Power Measurements
Polarization Dependency
Mating Stability of SOC Connector
Repeatability
Optical Interface Power Meter
Optical Interface MM Laser Power Source
1. While temperature is running profile from 0 °C to +50°C
Section 4: Specifications, Parts & Accessories
1
+10dBm to -65dBm
800nm to 1700nm
850nm, 980nm, 1300nm,
1480nm, 1310, 1550nm,
1625nm
< ±0.25dB (6%)
±0.22dB
±0.05dB
±0.22dB
<±0.05 dB (0°C - 50°C ±
1°C 24hr@>-30dBm)
0.5 second (typical)
dBm, dB, Watt
<0.10dB
<0.02dB
<0.05dB
SOC Adapter
UCI-UPC flat polish
Adapter, 62.5/125 MM
35
1

Advertisement