Electromagnetic compatibility
The AirMini complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC 60601-1-2:2014, for residential, commercial and light industry
environments. Portable and mobile RF communications equipment should be used no
closer to any part of the machine, including cables, than the recommended 10 cm
separation distance. The AirMini has been designed to meet EMC standards. However,
should you suspect that the machine performance (eg, pressure or flow) is affected by
other equipment, move the machine away from the possible cause of interference.
The AirMini complies with Part 15 of the FCC Rules and Industry Canada license-exempt
RSS standards. Operation is subject to the following two conditions: This machine may
not cause harmful interference, and this machine must accept any interference received,
including interference that may cause undesired operation.
FCC ID: QOQBT121, IC: 5123A-BGTBT121
Additional information regarding the FCC Rules and IC compliance for this machine can
be found on www.resmed.com/downloads/devices.
IEC 60601-1 (Edition 3.1) classification
Class II (double insulation), Type BF, Ingress protection IP22.
Maximum single fault steady pressure
Machine will shut down in the presence of a single fault if the steady state pressure
exceeds 30 cm H
O (30 hPa) for more than 6 sec or 40 cm H
2
1 sec.
Sound
Declared dual-number noise emission values in accordance with ISO 4871:1996
Pressure level measured according to
ISO 80601-2-70:2015 (CPAP mode)
Power level measured according to
ISO 80601-2-70:2015 (CPAP mode)
The A-weighted sound pressure level measured according to ISO 17510:2015:
AirMini tube connected to F20 connector for
AirMini with mask
AirMini tube connected to N20 connector for
AirMini or P10 connectors with mask
O (40 hPa) for more than
2
29 dBA with uncertainty of 2 dBA
37 dBA with uncertainty of 2 dBA
19 dBA with uncertainty of 3 dBA
15 dBA with uncertainty of 3 dBA