Page 3
PRECAUTIONS Prior to installation, operation or maintenance activities, carefully read this user manual. Follow the instructions strictly. Prior to the first use, carefully read this user manual. Use the device only as intended. Place loads in the centre of the weighing pan. Load the weighing pan with loads of gross weight which does not exceed the maximum capacity.
-10C to +40C. It communicates with various devices via the following interfaces: RS232, RS485, Ethernet, Profibus, and Modbus protocol. The TWM4 scale is equipped with 4 optoisolated inputs and 4 Opto MOS outputs. It is powered by 100-240VAC mains voltage.
Do not use cleansers containing abrasive substances. 4. SERVICE AND REPAIR In the case of any sign of damage, it is necessary to disconnect the device form the mains immediately. The damaged component must be replaced or repaired by RADWAG service directly.
5. RECYCLING TWM4 scales must be recycled, they are not to be treated as a regular household waste. Scales to be decommissioned must be decommissioned in accordance with valid legal regulations.
6.3. Scales with Load-Cell Modules TWM4 scale with load cell modules is intended to carry out mass measurement of silos. Modules are built into construction of ground-fixed silo's support. In most cases, the scale design includes 3-4 load cell modules. The modules, depending on the model, are made of stainless or galvanized steel.
PG11 cable gland for platform 1. PG11 cable gland for platform 2. PG11 cable gland for platform 3. PG11 cable gland for platform 4. M12 8P connector for RS232. M16 cable gland for Ethernet/RS485 cable (conditioned by the scale model). M16 cable gland for 4IN cable (option: PROFIBUS IN connector).
6.7. Inputs / Outputs 6.7.1. Technical Specifications Output parameters Output quantity Output type OptoMOS Maximum output current 0.2 A DC Maximum output voltage 50 V DC Input parameters Input quantity Input type Optoisolated Input voltage range 5 - 24 V DC 6.7.2.
Mind not to damage cable connecting the mass converter and a weighing platform. 7.1.1. 1-Load-Cell Plstform Scales of TWM4-x Series Take the scale out of the packaging. Place the scale on a flat and even surface. Keep it far away from any sources of heat.
Place the scale on an even surface, install ramps onto the steel brackets. Remove ramps, looking through bracket holes mark spots where openings for anchor bolts are to be made. Drill openings, mount steel brackets to the ground. 5 - nut 6 - washer 7 - centring cone...
7.3. Levelling: 4-Load-Cell Platform Scales To level the scale use the levelling feet and the level indicator. Each foot can be turned left and right, turning causes tilt. Adjustment span of the scale level is narrow therefore the correct level is obtained using steel washers which are to be put under the feet.
Keys: Press to zero the scale Press to tare the scale Press to select a weighing platform if the mass converter communicates with more than one platform. Zero and tare functions are available for the currently selected platform. 8. MwManager HOME SCREEN The procedure of communication establishing and a detailed description of connection between the scale and the ’MwManager’...
<ENTER> key on the keyboard. Buttons: MW-01 mass converter, TWM4 scale. MW-04 mass converter, TWM4 scale. Exit button. Press to close the program. The following message is displayed: <Close program?> (where: <Yes> - press to stop the program; <No> - press to go to the program menu back).
10.1. Good Weighing Practice: 1-Load-Cell Platform Scales In order to guarantee a long-term operation, wherein correct measurements are provided, the following principles must be adhered to: Load the weighing pan steadily, avoid applying mechanical shocks. Place the loads centrally on the weighing pan (eccentricity errors are specified by EN 45501 standard, point 3.6.2).
pallet scales (load placed on EUR-pallet loaded using pallet truck); middle block of pallet placed on the scale must remain unsupported: ramp scales (trucks used in meat plants); it is necessary to select platform matching the span of truck casters, this is to make sure that in the case of trucks of weight close to the maximum capacity, the load is transferred onto the platform close to load-carrying sections (profiles): 10.3.
I weighing range takes place automatically upon unloading of the weighing pan and returning to AUTOZERO – ( pictogram/marker is displayed). 11. TECHNICAL SPECIFICATIONS For technical specifications of respective scales go to RADWAG website www.radwag.pl.
12. ERROR MESSAGES Err2 Value out of zero range. Err3 Value out of tare range. Calibration mass or start mass beyond the acceptable range (1% for weight, 10 Err4 for start mass). Err8 Taring/zeroing time out of range. NULL Zero value from converter. FULL Weighing range exceeded.