Troubleshooting

When using the Vivatmo me, an error message may indicate a problem.
The following table gives explanations of the error messages as well as recommendations for solving the problem.
Display
Description
Regeneration or reconditioning process
is ongoing.
E-1
Permissible number of measurements
< 20.
E-2
Maximum number of permissible
measurements reached. Device does not
perform anymore measurements.
E-4
Humidity too high.
E-5
Ambient temperature or air pressure is
outside the specified range.
Solution
Wait for the process to be completed. Do not remove the
mouthpiece.
See display for number of remaining measurements.
Dispose Vivatmo me, see chapter "6.4 Disposal of device,
batteries and mouthpieces".
Check ambient conditions and take a new mouthpiece,
see chapter "2.1 Attaching the disposable mouthpiece".
Use the Vivatmo me in location with permissible ambient
conditions. When transporting the device from one place
to another, a stabilization period of min. 20 minutes
should be allowed before use.
Instructions for use
Maintenance and Troubleshooting