Programming Special Functions - Miele G 7881 Operating Instructions Manual

Dental thermal disinfector
Table of Contents

Advertisement

For future servicing, please
document any alterations to the
standard factory settings.
Enter any alterations in the field
Value set:________.
General notes
Program selector positions that have
not been allocated a function are
indicated in the middle part of the
display by a bar (-).
The programming level is called up via
the "Stop" f position.
To program special functions
Enter the programming level of the
controls first before performing special
functions as follows:
1. Setting the liquid detergent
dosage and priming the
dispensing system
^ Turn the program selector to the
"Stop" f position.
^ Turn off the machine.
^ Press and hold "Display" ß and
"Start" 6 at the same time.
Simultaneously, turn the machine on
with the "On/Off" button.
The current program status "P..." will
appear in the display and the
"Fill/Drain" indicator will illuminate.
^ Press "Drying" 0 twice.
"E02" (programming level) wil appear
in the display.
Value set:

Programming special functions

1.1 DOS 1 (liquid detergent) via the
optional DOS-Module G 60
1.1.1 Setting the dosage
Follow the dosage recommedation
of the cleaning agents manufacturer.
^ Turn the program selector to "Wash".
The display will show either of the
following:
Display
10
0.10 - 1.00
^ To change the amount of liquid
detergent dispensed, press "Display"
ß as many times as necessary to
achieve the desired result (see chart
below):
Setting
.10
.20
.30
.40
.50
.60
.70
.80
.90
1.0
* Recommended dosage for neodisher
Dosing
Powdered detergent
from cup
Liquid detergent from
DOS-Module G60
Dosage
20 ml
30 ml
40 ml
50 ml*
60 ml
70 ml
80 ml
90 ml
100 ml
110 ml
TM
FA
37