Priming the neutralizer
dispensing system
^ Turn the program selector to the
"STOP" f position.
^ Turn off the machine.
^ Press and hold the "Display" ß and
"Start" 6 buttons at the same time.
Simultaneously turn the machine on
with the "On/Off" button.
The current program status "P..." will
appear in the display and the
"Fill/Drain" indicator will illuminate.
^ Press the "Drying" 0 button twice,
"E02" (programming level) will appear
in the display.
^ Turn the program selector to the
5 o'clock position.
"do3" appears in the display.
^ Press "Display" ß .
The dispensing system will be
primed automatically. "do3" will
appear in the display when priming is
finished.
Every time the dispensing system has
been primed:
^ Run the "Rinse" program to dilute and
drain away any agents that may
remain in the wash cabinet.
Adding neutralizing agent
Refilling the neutralizing agent
When the "Neutralizer" indicator
illuminates:
^ Replace the current neutralizer
container with a full one.
^ The indicator light should go out.
^ If the container has been allowed to
run dry, the light will not go out until
the priming process is performed.