Other Potential Failure Modes - Carrier UG 15 Tier 4 Operation And Service Manual

Engine emission system, generator sets
Table of Contents

Advertisement

Alarm
Name
38
System Over Temp
44
Filter Outlet Temp
45
Exhaust Inlet Temp
53
ECM Output Failure
54
ECM Log Failure
55
ECM Clock Failure
62
ECM Temp Low
63
ECM Temp High
5.2

Other Potential Failure Modes

Cracked filter substrate
Blocked Doser, lack of fuel supply, Doser problems
Ruptured DPF housing or broken exhaust piping
Excessive fuel and lube oil consumption
Cause / Troubleshooting
Filter Outlet Temperature is more than 850°C during regeneration.
In some rare cases regeneration may produce higher than expected
temperatures leaving the DPF. This may be just after repairs with the back
pressure still high. This indicates that there is excess soot in the DPF. In these
cases, the EES will try to reduce temperatures by stopping fueling.
If temperatures still exceed desired values, the unit will temporarily shut down
to limit the burn rate of the soot.
After about 15 minutes the unit will restart and regeneration will continue if
needed.
Filter Outlet Temperature is less than -100°C or more than 1100°C at any
time.
Inspect the Filter Outlet Temperature Sensor (ESFOT). Refer to
3.6.3.
Check for exhaust leaks as this may indicate a leak in the system.
Replace if necessary.
Exhaust Inlet Temperature is less than -100°C or more than 1100°C at any
time.
Inspect the Exhaust Inlet Temperature Sensor (ESEIT). Refer to
3.6.1.
Replace if necessary.
ECM DO Current above 32A or output shorted to positive or negative supply
or DO output circuits temperature above 125°C for more than two seconds.
Check fuse F15.
Perform a manual regeneration. Refer to
Replace the ECM if no outputs become energized.
ECM is not able to log events.
ECM internal data logger has failed.
Replace the ECM.
ECM is not able to initialize.
ECM internal clock has failed.
Replace the ECM.
ECM Temperature is less than -43°C.
ECM internal temperature is below the accuracy threshold for the device.
Allow the ECM to warm up.
If alarm is not cleared after ECM is warmed above -38°C, replace the ECM.
ECM Temperature is more than 125°C.
ECM internal temperature is above the accuracy threshold for the device.
Allow the ECM to cool.
If alarm is not cleared after ECM is cooled below 120°C, replace the ECM.
5–3
Section
Section
Section
4.3.
T-378

Advertisement

Table of Contents
loading

Table of Contents