Adjustment Instructions; Step-By-Step Instructions - GE TONOPORT V Service Manual

Hide thumbs Also See for TONOPORT V:
Table of Contents

Advertisement

10

10.1 Step-by-Step Instructions

Revision H

Adjustment Instructions

Only the pressure amplifier needs to be calibrated. This is done by switching on
the TONOPORT V and waiting until the time is displayed on the LCD.
Measure the direct voltages at TP2 and TP3. This is done by turning the
trimming potentiometer R59 with a screwdriver until the voltages at TP2 is 1.8
±0.1 V and at TP3, 2.0 ±0.1 V.
Press "INFO" four times, LCD indicates "H 4".
Press "START" once, a 2-digit figure is displayed on the LCD.
Use a screw driver to turn the trimming potentiometer R53 until the LCD
indicates a value of approx. 50.
Connect a reference pressure gauge, a rubber bulb and air container (1 l) to the
TONOPORT pressure connector and press "START".
Using the rubber bulb generate a pressure of 100 mmHg in the tubing system
and watch the LCD.
Using a screwdriver, turn the trimming potentiometer R40 until the LCD
indicates a value of 100.
Also check the values indicated on the LCD at pressures of 50 mmHg, 150
mmHg and 200 mmHg. Max. permitted deviation: ±3 mm Hg.
Switch the TONOPORT off.
TONOPORT V
2001589-023
10-1