Comark RF400 Help Guide – Troubleshooting

Forced Connection and Transmission

It's possible to get the RF400 Logger to reconnect to a network and upload any unsent readings to the Comark
Cloud. To do this hold down the button on the front of the RF400 Logger and it will beep once. Keep the button
pressed until the RF400 Logger beeps again then release the button. This will force a connection and transmission.

Signal Strength

The RF400 Logger reports its current signal strength to the Comark Cloud
every time it communicates. It is possible to get real time signal strength
from the RF400 Logger to aid installation issues. To do this hold down the
button on the front the RF400 Logger and it will beep once. Keep the button
pressed until the RF400 Logger beeps two more times, then release the
button. The RF400 Logger will start beeping in relation to the signal strength.
Fast beeping is a good signal, slow beeping is a poor signal. In addition, the
LED on the button will flash green indicating a good signal, orange indicating
a medium signal and red for a poor signal. Press the button once more to
end signal test. Signal strength will only work if the RF400 Logger has been
set up to talk to the Comark Cloud. Signal strength test will automatically
end after 1 minute.
LED & Buzzer Patterns During Normal Operation
Quick Green LED every 5 seconds
Solid Green LED
Red LED every 5 seconds
Quick Red LED every 5 seconds
Quick Orange LED
(short beep every 30 min)
Solid Orange LED
(short beep every 30 min)
Comark Instruments
52 Hurricane Way, Norwich, Norfolk, NR6 6JB, UK
Tel: +44 207 942 0712
www.comarkinstruments.com
Comark Instruments
PO Box 500, Beaverton, OR 97077, USA
Tel (503) 643 5204
www.comarkUSA.com
RF400 Loggers Trouble Shooting Guide
Issue 1 – Oct 2019
Normal operation – battery power
Normal operation – external power
Alarm Active
Alarm mute
Generic Fault e.g. Lost network connection or missing probe – battery
power
Generic Fault e.g. Lost network connection or missing probe – external
power
The Professionals' Choice

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the COMARK RF400 and is the answer not in the manual?

Questions and answers

Summary of Contents for Fluke COMARK RF400

  • Page 1 RF400 Loggers Trouble Shooting Guide Issue 1 – Oct 2019 Comark RF400 Comark RF400 Help Guide – Troubleshooting Forced Connection and Transmission It’s possible to get the RF400 Logger to reconnect to a network and upload any unsent readings to the Comark Cloud.
  • Page 2 LED & Buzzer Patterns During Setup Orange LED blinks once every 2 seconds RF400 is in Setup and waiting for an App to connect to it Orange LED blinks twice every 2 seconds The App has connected to the RF400 and is waiting for setup information (quick beep) Setup in progress.
  • Page 3 deleting all settings and clearing any remaining data. The device will no longer be connected to any network or Comark Cloud account. IT IS RECOMMENDED THAT YOU ARCHIVE THE DEVICE ON YOUR COMARK CLOUD DEVICES PAGE BEFORE UNDERTAKING A FACTORY RESET. How do I turn my RF400 LOGGER device off? The device will remain on permanently unless the battery runs out.
  • Page 4 If my RF400 LOGGER device's battery runs out, will I lose all of my data? The device will not take readings when there is no power but all readings it has taken prior to the loss of power will be synchronised. Once the batteries have been replaced, the device will reconnect to the Comark Cloud, start a new recording session and continue recording.

Table of Contents