Set Up Of The Liquid Drain; Connecting The Unit To The Mains - Elma Elmasonic Denta Pro Operating Instructions Manual

Table of Contents

Advertisement

Initial operation
5.1
Proceed as follows
5.2
Required mains
conditions
Connect mains cable
12

Set up of the liquid drain

On the delivered unit, the drain duct for the cleaning liquid is
closed off with a plastic screw cap.
For setting up the liquid drain fix the delivered tube socket to
the drain duct.
1. Unscrew (clockwise) the plastic screw cap (see illustration
5.1)
2. Screw the tube socket (included in delivery) onto the inside
thread of the drain duct (clockwise).
3. Turn the tube socket into the required drain position (see
illustration 5.2).
The plastic thread is self-sealing when the socket has been
screwed in by hand as far as possible.
Note: Unscrewing the tube socket (anti-clockwise) can
cause a leak of the thread.
4. The drain duct is now ready for connection to a customer-
provided discharge system. Use a standard tube (dia 1/2").
Push the tube onto the socket and fix it with the clamp
included in the delivery.
Illustration 5.1
Drain with plastic screw cap

Connecting the unit to the mains

Earth grounded socket:
1 phase (220-240 V); 1 N; 1 PE protective earth.
The power supply must be protected by an earth leakage circuit
breaker.
Use the plug-in mains cable delivered with the unit. Connect the
unit to a grounded shockproof socket only. Ensure that the
values indicated on the nameplate of the unit must correspond
with the available connecting conditions.
The mains plug must be connected to an easily accessible
socket only, as it serves as interrupted device!
Illustration 5.2
Drain fitted with standard tube
© Elma Schmidbauer GmbH

Advertisement

Table of Contents
loading

This manual is also suitable for:

Elmasonic denta pro h

Table of Contents