7.
Specifications
7.1
Probe
Type
Acoustic working
frequency
Technology
Dimensions
Weight
Probe Material
Acoustic output
Applicable system
Classification
Cleaning
Disinfection
Sterilization
Operating conditions
Ambient temperature
Contact surface temperature
(temperature of examinee)
Relative humidity
Storage conditions:
Temperature
Relative humidity
: L34 Probe
: 5.0MHz
: High density linear array probe
: See
Fig. 8
: Approx. 0.44kg
(incl. cable and connector)
: Bio-compatible allergy free components
: According to IEC60601-2-37
(See Main Unit manual.)
: Depending on production and upgrade
status. For detailed information contact
a service support.
: MDD classification Ⅱa.
: Applicable detergents are listed in
the suppliers list.
: Applicable disinfectants are listed in
the suppliers list.
: ETO gas sterilization
Plasma sterilization
:
; +10 – +35°C
; max. 42°C
; 30 – 85%
; –10 – +55°C
; 10 – 95%
(subject to no condensation)
-19-
Q1E-EP1361