Hitachi L34 Instruction Manual page 19

Hide thumbs Also See for L34:
Table of Contents

Advertisement

Manual disinfection:
1) Prepare the disinfectant solution in a tank with cold water (please
follow the instructions of the disinfectant manufacturer regarding
application, concentration, microbiological efficiency, service
life and contact time).
2) Confirm the concentration of the disinfectant before immersing the
probe. Although Cidex® OPA does not need to be diluted, it is
recommended to use test strips to verify the concentration. The test
strips can indicate whether or not the concentration is above the
Minimum
Effective
expiration date of the test stripes. Temperature of disinfectant
solution should be minimum 20 °C. The minimum contact time is 5 minutes.
If a different disinfectant is used, follow the manufacturer's
instructions. Please also consider the material compatibility for the
medical device.
3) Immerse the immersible part of the probe into the disinfectant (see
). Set a clock to insure the recommended contact time which is
Fig. 4
5 minutes.
4) Rinse the immersible part of the probe with deionized water for 1 minute.
(alternatively: immerse the immersible part of the probe in a tray
filled with deionized water (see
5) Visually check the outer surface of the probe for leavings of the
disinfectant. If necessary, repeat the rinsing.
Concentration
Water
Detergent
Disinfectant
Fig. 4 Immersion of the probe
(MEC).
Please
) for 5 min.)
Fig. 4
-13-
also
note
the
Q1E-EP1361

Advertisement

Table of Contents
loading

Table of Contents