Reprocessing Procedure

The probe must be reprocessed after each use. Refer to the
reprocessing instruction in this chapter. For reprocessing of
Coupler attachment EZU-PA3C1, PA3C1H and PA3C2, refer to the
instruction manual of each Coupler attachment.
WARNINGS
Limitations on
reprocessing
Transportation
before using
The probe is delivered unsterile. Prior to the first
use, reprocess the probe.
Temperature should not exceed 60°C during
reprocessing.
Probe connector is not water resistant.
The probe is not completely submersible. The immersible
part is shown in Fig.1. The un-immersible part should
be disinfected by wipe disinfection.
The probe should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful not
to damage the sterile pouch or container during
transportation.