3.5.5 Protocol set-up GE Medical System Senographe

Select Protocol (Shift-F1) when the SENOGRAP protocol is selected. This will bring up
the protocol set-up for the GE Medical System Senographe.
Define text for GE Senographe
Mode
Sub Mode
Focus
Track
Filter
This window is used to define a number of text strings.The GE Medical System
Senographe sends a few parameters as 0 and 1, whereas they should be presented as
texts instead. This window is used to define those texts. For more detailed information,
please refer to the GE Medical System Senographe manual.
Each string may be up to 8 characters long. Use the up and down arrows to go to the row
below or above, and use Left- and Right soft-keys to select column.
Creating the picture that will be printed on the film is done in the same way as usual
selecting the COMM field type. NOTE! The communication protocol has to be set before
setting up the picture, otherwise the COMM field will display NONE.
Example
We want to create a COMM field that shall display the MODE. Create a new field and
under fielddefinition select Type COMM. The The Senographe is sending 0, 1 or 2 as the
MODE parameter and the Kodak Network ID Camera is translating it to a text according
to the protocol set-up above. In this case the texts are AOP for 0, AEC for 1 and Manual
for 2. With reference to this settings, the length should be set to 6 and Comm to MODE.
This means that we have to define the length of this COMM field to six characters
maximum. Select Comm MODE will link the Mode data to this field.
Fielddefinition
Type: COMM
Row:
Comm: MODE
The field data do not include the unit (kV, mAs etc.). To add that, an F-text field with the
appropriate text has to be inserted after the COMM field. Create a new field and under
fielddefinition select Type F-text. Enter values for the row and column to place this field at
a position of your preference. Hint, use the Place function to position fields, see chapter
3.4.3. Enter appropriate text under Fixed text.
Fielddefinition
Type: F-text
Row:
Fixed text:
Mode
Kodak AG Stuttgart
AOP
AEC
STD
CTS
SF
LF
RH
MO
RH
AL
Left
Right
Length: 6
56
Column: 108
Font: Small
Previous
Next
56
Column: 84
Font: Small
Previous
Next
Manual
DOSE
MO
Save
Save
Save
15
Installation and Set-up Manual