Hi30033 Probe - Hanna Instruments Groline HI981413 Instruction Manual

Nutrient dosing system
Table of Contents

Advertisement

• High & Low with enable or disable option
• Triggered after 5 sec. if the controller records consecutive readings over or
EC/TDS Alarms
under threshold values
• Level with enable or disable option
• Overtime protection (1 to 180 min. or Off)
Controller
• Intuitive alarm system, using red, yellow and green color coded backlight
Alarm System
• User-selectable, alarm setup options
• SPDT 2.5A / 230 Vac
Alarm Relay Output
• Activated by EC/TDS - selectable alarm conditions
• Quick connect DIN connector
Probe Input
• Galvanic isolation
• External switches can be attached to stop the pump and activate the alarm
Level Sensor Input
(low reagent level) when the switch is open and level alarm is configured
(Digital Input)
in Setup
• Galvanic isolation
Power Supply
100 - 240 Vac, 50/60 Hz
Power Consumption 15 VA
Environment
0 to 50 °C (32 to 122 °F), max. 95% RH non-condensing
Dimensions
90 x 142 x 80 mm (3.5 x 5.6 x 1.8")
Weight
908 g (36 oz)
Casing
Wall mounted, built-in pump, IP65 rating

4.2. HI30033 PROBE

Range
Pin
Temperature sensor
Temperature range
Body
Top thread
In-line mounting thread
Cable length
Connector
Maximum pressure @25 °C
0-10 mS/cm
Stainless steel
Yes
-5 to 60 °C (23 to 140 °F)
PVDF (white)
3/4" NPT
1/2" NPT
2 m
Quick connect DIN connector
3 bar (43.5 psi)
7