SPECIFICATIONS continued
TEMPERATURE SENSOR TYPE:
SETTING
200 °C
630 °C
*ACCURACY:
REPEATABILITY:
TEMPERATURE COEFFICIENT:
MAXI
MUM LEAD RESISTANCE:
(each lead):
SENSOR CURRENT:
MODELS 868 AND 869
GENERAL SPECIFICATIONS:
DISPLAY:
CONVERSION RATE:
OVERRANGE AND OPEN
SENSOR INDICATION:
MODEL 869
RANGE
-100.0 °C to 199.9 °C
-199.9 °C to -100.1 °C
-100 °C to 630 °C
-220 °C to -101 °C
Three wire accuracy is the same if contact
resistance errors are removed by calibration
of instrument plus probe at 0°C. Includes DIN
43760 (ITS-90) conformity, repeatability,
temperature coe cient (18° to 28 °C), time
stability (one year) and errors with up to 50 Ω
of lead resistance (each lead). Excludes probe
errors; however, probe errors around 0°C may
be compensated by an internal adjustment.
0.1 °C
Typical for one week at constant
ambient temperature.
Four wire: 50 Ω
Three wire: 10 Ω
500 µ A max.
3
/
digit LCD, 0.5" (13 mm) height. Polarity and deci-
1
2
mal point indication
1.5 readings per second.
Three wire or four wire 100 Ω
platinum RTD (alpha = .00385)
RESOLUTION
0.1
0.1
1
1
18° to 28°C; included in
accuracy speci cation. From
-10° to 18°C, and 28° to
50°C: less than ±0.015°C/°C.
.
22
4 WIRE
ACCURACY*
(18 °C TO 28 °C: 1 Year)
±
0.3 °C
±
1.5 °C
±
1 °C
±
2 °C