palintest Lumiso Chlorine User Manual page 21

Table of Contents

Advertisement

8
LMP001
Blank
Lumiso
Lumiso
A Palintest Product
A Palintest Product
12:28:49
16-07-2021
Pooltest
Should any standard read outside the values shown
on your certificate, the most likely cause is contamination
on the optical windows in the cell holder. Please clean
the cell holder carefully and remeasure the standards.
(See 'Keeping the Optics Clean')
If this does not resolve the issue, please contact your local
Palintest branch or distributor.
530
nm
Continue, by inserting all the
remaining standards, pressing
read, and comparing the result
with certificate.
21