7.2 Position accuracy of the sensor

The following illustrations show the tolerance margins of the sensor position relative to the outer camera
housing. A maximum error in all directions (rotation, translation) cannot occur at the same time.
Fig. 3: Position accuracy of the sensor (1)
It cannot be guaranteed that the sensor cover glass or filter glass are absolutely parallel to each other.
C-mount lenses can also be subject to inaccuracies of the flange back distance. The
tolerance usually is ±0.05 mm. In some cases, however, the inaccuracies of camera and lens
might add up, resulting in a total error > 0.05 mm.
Position accuracy, in each direction
Horizontal/vertical rotation (α)
Translational rotation (β)
Flange back distance
January 2020
Technical Manual: GigE uEye CP Rev. 2
Fig. 4: Position accuracy of the
sensor (2)
±0.3 mm
±1.0°
±1.0°
±0.05 mm