Questions/Reminders
Options:
Off or Questions 1 – 51(Classic mode only);
Questions 1 – 61(Genesis DM with DSSM activated)
Default:
Off
Access Level:
General
As you scroll through the block of questions and/or reminders, you see blinking cue text and the
question ID number. The ID number is located in the upper right-hand corner of the display. Each
question has its own unique ID number that separates it from the other questions in the list.
Each Question has an ID number and an associated question. For example, the prompt cue text (seen
by patient) reads. "Out of Medications?" This prompts' ID is 07. It is the seventh question of the 51
total questions. The #1 designates the place order in which the question will be presented to the
patient. So in this example Question 7 of 51 is being presented first.
Question #1
07
Out of medications?
For the list of all questions, refer to Appendix A Subjective Questions.
Question Time
Options:
OFF, every time for or any combination of days ( _ _ Mo _ _ We _ _ Fr _ _), or only one
time each day ( _ _ Mo* _ _ We* _ _ Fr* _ _ )
Default:
All days
Access Level:
General
Each question is followed by an option that allows you to set a question frequency. The default is daily.
However, you can program any combination of days. You must have at least one day selected or the
monitor will not ask the question.
To program a question and question time
1.
In Questions Menu, scroll to the desired question, and press RETEST. The Off text (or cue text)
blinks.
2.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
3.
Press YES to scroll to the question time, and press RETEST to enter. The first selection blinks.
4.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
5.
Continue through the seven days until all are set correctly.
6.
Scroll to the next question you want to program and repeat steps 1 – 5.