Side Ports; Lithium Battery; Other System Equipment - Honeywell GENESIS DM Clinicians Manual

Hide thumbs Also See for GENESIS DM:
Table of Contents

Advertisement

GENESIS DM CLINICIAN GUIDE
Equipment Overview
(4) Isolated serial I/O ports (SCALE, COM1, COM2)
The three isolated ports with Mini-DIN 4 connectors supply up to +5VDC @ 250mA. The first
isolated port is a dedicated scale COM port. COM1 and COM2 allow a patient to use an external
device connected to the monitor. Isolation protects patients from any potential of shock.
Connector cables for external devices only fit the port appropriate for the device.

Side ports

The following figure shows the ports located on the side of the Genesis DM monitor.
(1) Genesis DM SmartMedia Card slot
The card slot holds a SmartMedia Flash Memory card that stores voice and text prompts and
customizable patient questions.
Sentry and previously released Genesis cards ARE NOT compatible with the Genesis DM.
(2) Non-invasive blood pressure (NIBP) pneumatic port
The NIBP pneumatic port with its quick-release connector is the attachment site for the blood pressure
hose.

Lithium battery

The lithium battery maintains the memory, clock and configuration for approximately five years. You
cannot replace the lithium battery. If the battery fails, contact Customer Service to have it
replaced.

Other system equipment

In addition to the monitor, the monitor includes a scale with a scale cord, blood pressure cuff and hose,
power supply, and a power strip. The SpO
finger sensor with external oximeter module cable is an
2
optional accessory. This section provides a general overview of each piece of equipment. For detailed
specifications, refer to Appendix D, Equipment Specifications.
12/5/2007
12
P.4510EN.01

Advertisement

Table of Contents
loading

Table of Contents