You can scroll to next System Settings option -or-press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Language
Options
(DM Only):
American English – Male and Female (clinician selectable)
Default:
American English - Female
Access Level:
General
The Language menu allows you to change the language of the audio and text prompts. To change the
language setting, follow the steps below.
1.
In the System Settings menu, scroll to Language, and press RETEST to enter. The current setting
blinks.
2.
Press either YES or NO to scroll to the desired language, and press RETEST to enter the setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.
Units
English (Lbs and °F) or Metric (Kg and °C)
Options:
Default:
Set to common usage for your locale
Access Level:
General
You can set the Genesis DM to display either English or metric units. Genesis DM monitors should
default to what is customary in your locale. However, if you need to change the units, follow the steps
below.
1.
In the System Settings menu, scroll to Units, and press RETEST. The current selection blinks.
2.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.
Inflation Settings
Options:
50 – 260 mmHg
Default:
175 mmHg
P.4510EN.01
GENESIS DM CLINICIAN GUIDE
Programming the Genesis DM Monitor
12/5/2007
25