1 0 . S U P E R V I S O R M A I N M E N U
10.4.6 Automatic shutdown
The SoToxa™ Mobile Analyzer can be set to automatically shutdown after a period of inactivity to save power. The automatic shutdown
function can be set to on or off and the time period before shutdown can be selected.
Select Analyzer Settings from the supervisor menu, then from the Analyser Settings menu, select Auto Shutdown. From the Auto Shutdown
menu, use the five way navigation key to highlight the required setting and press OK to save the setting.
The analyzer will not automatically shutdown while a test is in progress or if a SoToxa™ Test Cartridge is inserted.
10.5
Installed cartridges and updating cartridge types
To use a drug test cartridge on the SoToxa™ Mobile Analyzer, the test cartridge type must be installed in order that the analyzer recognises the
test cartridge. New test cartridge types can be added to the analyzer via the SD card.
To add a new test cartridge type using the SD card, follow these steps:
Step 1: Insert the SD card into the analyzer before starting this procedure.
Step 2: From the Supervisor menu, select the Installed Cartridges option.
Step 3: In the Installed Cartridges screen, press Add, on the bottom left of the screen.
Step 4: The analyzer will display the list of files and folders carried on the inserted SD card. To open a folder, select the folder, then press the OK
button. To exit a folder, select the top folder marked with an arrow, then press the OK button.
Step 5: Select the desired test cartridge file and press the Enter button.
33 | EN