Advanced Meter Functions - Fluke 123 Service Manual

Industrial scopemeter
Hide thumbs Also See for 123:
Table of Contents

Advertisement

Open Circuit Voltage
Continuity (CONT)
Beep
Measurement Current
Detection of shorts of
Diode
Maximum Voltage:
@0.5 mA
@open circuit
Accuracy
Measurement Current
Polarity
Capacitance (CAP)
Ranges
Accuracy
Full Scale Reading
Measurement Current
Measurement principle

2.3.3 Advanced Meter Functions

Zero Set
Fast/Normal/Smooth
Meter settling time Fast
Meter settling time Normal
Meter settling time Smooth
Touch Hold (on A)
TrendPlot
Fixed Decimal Point
<4V
30Ω ± 5Ω in 50Ω range
0.5 mA
≥1 ms
>2.8V
<4V
±(2% +5 counts)
0.5 mA
+ on input A, - on COM
50 nF, 500 nF, 5 μF, 50 μF, 500 μF
±(2% +10 counts)
5000 counts
5 μA to 0.5 mA, increases with increasing
ranges
Dual slope integrating measurement with
parasitic serial and parallel resistance
cancellation.
Set actual value to reference
1s @ 1μs to 10 ms/div
2s @ 1μs to 10 ms/div
10s @ 1μs to 10 ms/div
Captures and freezes a stable measurement
result. Beeps when stable. Touch Hold
works on the main meter reading , with
threshholds of 1 Vpp for AC signals and
100mV for DC signals.
Graphs meter readings of the Min and Max
values from 15 s/div (120 seconds) to 2
days/div (16 days) with time and date
stamp. Automatic vertical scaling and time
compression.
Displays the actual and Minimum,
Maximum, or average (AVG) reading.
Possible by using attenuation keys.
Characteristics
2
2.3 Dual Input Meter
2-9