Chapter 7 Instrument Maintenance and Troubleshooting
7.1 Daily Use and Maintenance of Instrument
7.1.1 Connector Cleaning
This series of OTDR optical output interface for the replacement of the universal optical
interface, the use of the process must be maintained in the end of the cleaning. When the
instrument is unable to test the normal curve, or the test result is not accurate, the connector
should be considered first.
When cleaning, make sure that the OTDR and the visual red fault location function are
stopped. Unscrew the output interface, use a special paper towel or cotton swab moistened with
alcohol to wipe the connecting end face.
At the same time, in the normal use of the instrument, after the completion of the dust cap