Reference Guide
Replacing air temperature and humidity
sensor
•
Electric joint compound
•
Screwdriver Torx 10
1
1.
Loosen the 4 screws that hold the temperature sensor in place, pull the air temperature and
humidity sensor (HMP) through the holder, and disconnect the sensor cable.
2.
Add electric joint compound at the end of the connector of the replacement sensor.
Connect the sensor cable to the replacement sensor.
3.
Remove the yellow cover from the replacement sensor.
4.
Push the replacement sensor through the holder and attach the holder with 4 screws.
More information
‣ MD30 maintenance (page 1)
‣ MD30 maintenance checklist (page 15)
6